| dc.description.abstract | RESUMEN :
INTRODUCCIÓN: La capacidad de presurización de los dispositivos de presión positiva continua
dependen de factores intrínsecos y extrínsecos; sin embargo, no se encuentran estandarizados,
generando errores en la elección de dispositivos frente al aumento de la demanda ventilatoria,
especialmente en pacientes con patologías respiratoria.
OBJETIVOS: Determinar las diferencias de presurización entre los dispositivos CPAP con aumento
de la demanda inspiratoria, analizando la concordancia entre la presión prescrita y la presión dada
entre los modelos.
DISEÑO: Evaluación de 4 ventiladores (Yuwel, V60-Philips, PB560-Medtronic y Lumis150-Resmed)
en un estudio piloto simulado, con dos patrones respiratorios y tres esfuerzos inspiratorios,
incluyendo tres niveles de fuga, y considerando dos presiones pautadas para valorar la capacidad
de presurización.
MEDICIONES PRINCIPALES: Para el diseño descrito, se analizaron 768 ciclos respiratorios. La
capacidad de presurización se evaluó mediante las variables “height” (cmH2O), “Width” (ms) y “peak
área” cmH2O/seg correspondiendo a la altura en negativa del pico de presión, el tiempo desde inicio
a fin del esfuerzo y el área formada por ambas, respectivamente.
RESULTADOS: La peor presurización correspondió a PB560, mientras que la mejor presurización
correspondió a Lumis 150 y V60 (p<0.001 ANOVA test). El ventilador V60 obtuvo la mejor capacidad
de adaptación. La presión prescrita más exacta fue V60 con media de diferencia de presión −0.19
cmH2O±0.06 en modalidad CPAP a presiones pautadas de 8 y 12 cmH2O.
CONCLUSIONES: Existen diferencias estadísticamente significativas entre los modelos evaluados
en cuanto a su capacidad de presurizar frente a los distintos eventos que pueden producirse en la
práctica clínica.
PALABRAS CLAVE : “Continuous Positive Airway Pressure”, “Computer Simulation”, “Noninvasive Ventilation”, “Sleep Apnea Syndromes” | es_ES |
| dc.description.abstract | ABSTRACT :
INTRODUCTION: The pressurization capacity of continuous positive airway pressure devices
depends on intrinsic and extrinsic factors; however, they are not standardized, generating errors in
the choice of devices when increasing the ventilatory demand, especially in patients with respiratory
pathologies.
OBJECTIVES: To determine the differences in pressurization between CPAP devices with increased
inspiratory demand, analyzing the concordance between the prescribed and the given pressure
between the models.
DESIGN: Evaluation of 4 ventilators (Yuwel, V60-Philips, PB560-Medtronic, and Lumis150-Resmed)
in a simulated pilot study, with two respiratory patterns and three inspiratory efforts, including three
leak levels, and considering two prescribed pressures to assess the pressurization capacity.
MAIN MEASUREMENTS: For the described design, the study analyzed 768 respiratory cycles. The
variables "height" (cmH2O), "Width" (ms), and "peak area" (cmH2O/sec) corresponding to the
negative height of the pressure peak, the time from the beginning to the end of the effort, and the
area formed by both, respectively, were used to evaluate the pressurization capacity.
RESULTS: The worst pressurization corresponded to PB560, while the best pressurization
corresponded to Lumis 150 and V60 (p<0.001 ANOVA test). The V60 ventilator got the best
adaptability. The most accurate prescribed pressure was V60 with a mean pressure difference of
−0.19 cmH2O±0.06 in CPAP mode at prescribed pressures of 8 and 12 cmH2O.
CONCLUSIONS: There are statistically significant differences in the models evaluated in their ability
to pressurize against the different events that can occur in clinical practice.
KEYWORDS : “Continuous Positive Airway Pressure”, “Computer Simulation”, “Noninvasive Ventilation”, “Sleep Apnea Syndromes” | es_ES |
| dc.format.extent | 87 p. | es_ES |
| dc.language.iso | spa | es_ES |
| dc.rights | Attribution-NonCommercial-NoDerivatives 4.0 International | es_ES |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | es_ES |
| dc.subject | Teràpia respiratòria--Aparells i instruments | es_ES |
| dc.subject.other | Respiratory therapy--Equipment and supplies | es_ES |
| dc.title | Influencia del modelo de dispositivo de presión positiva continua sobre la presurización en patologías respiratorias en un modelo de banco de pruebas | es_ES |
| dc.type | masterThesis | es_ES |
| dc.description.degreename | Màster Universitari en Fisioterapia del Tòrax | es_ES |
| dc.rights.accessRights | openAccess | es_ES |